Standard curve calculator

Share this page

Favorites

    No favorites yet. Press Favorite on any page to keep it here.

    Methodology

    This standard curve calculator fits a straight line through your standards and reads unknown samples off it. Enter the known concentrations with their signals (absorbance, peak area, fluorescence), then the signals of the unknowns, and it returns the equation, R² and each concentration.

    Standards
    Concentration (x)Signal (y)Remove

    Separate values with commas, spaces or new lines.

    1 if the samples were not diluted.

    Standard curve y = 0.0495571x + 0.00338095
    Slope0.0495571
    Intercept0.00338095
    R²0.999784
    Standards used6
    SignalConcentration from the curve× dilution (1)Range check
    0.254.976464.97646Within the standards
    0.3627.236487.23648Within the standards
    Show the working, step by step
    1. Fit the least-squares line through the standards.

      slope = Σ(x − x̄)(y − ȳ) ÷ Σ(x − x̄)² = 3.469 ÷ 70 = 0.0495571 intercept = ȳ − slope × x̄ = 0.251167 − 0.0495571 × 5 = 0.00338095

    2. Rearrange the line to give concentration from signal.

      x = (y − intercept) ÷ slope

    3. Unknown with signal 0.25.

      x = (0.25 − 0.00338095) ÷ 0.0495571 = 4.97646

    4. Unknown with signal 0.362.

      x = (0.362 − 0.00338095) ÷ 0.0495571 = 7.23648

    The standard curve formula

    signal = slope × concentration + intercept concentration = (signal − intercept) ÷ slope

    The slope and intercept come from a least-squares fit of signal on concentration, the same fit the linear regression calculator performs. Reading an unknown is the line run backwards: subtract the intercept from its signal and divide by the slope.

    Worked example

    Six standards measured by absorbance:

    Concentration (mg/L)Absorbance
    00.002
    20.105
    40.198
    60.304
    80.401
    100.497
    1. Fit the line: y = 0.0495571x + 0.00338095, with R² = 0.999784.
    2. An unknown reads 0.250. Concentration = (0.250 − 0.00338095) ÷ 0.0495571 = 4.976 mg/L.
    3. A second unknown reads 0.362. Concentration = (0.362 − 0.00338095) ÷ 0.0495571 = 7.236 mg/L.
    4. If the samples had been diluted 1 in 5, multiply each result by 5.

    Both signals sit between the lowest standard (0.002) and the highest (0.497), so both readings are interpolations and can be reported.

    Checks before you trust the curve

    • R². Compare it with the limit in your method. The coefficient of determination page explains what it measures.
    • Residuals. Standards that fall above the line at both ends and below it in the middle show curvature that R² hides. The residual calculator lists each one.
    • Range. Report only unknowns whose signal lies between the lowest and highest standard.
    • The blank. An intercept far from zero points to a contaminated blank or a background signal that was not subtracted.
    • Replicates. Measure unknowns more than once and report the spread as a relative standard deviation.

    Where standard curves are used

    • Spectrophotometry. Absorbance against concentration, following the Beer-Lambert law.
    • Protein assays. Bradford and BCA assays read protein against a BSA standard series.
    • Chromatography. Peak area against the concentration of a reference standard.
    • ELISA and qPCR. These often need a curved or log-scale fit. A straight line is only right over the part of the range where the response is linear.

    This calculator fits a straight line only. If your standards bend, restrict them to the linear part of the range or use a curved fit such as the quadratic regression calculator.

    Standard curve calculator: the worked example on this page, with its result and chart
    Standard curve calculator: the worked example above, at a glance.

    Common questions

    What is a standard curve?

    A standard curve, also called a calibration curve, is a plot of instrument signal against known concentrations. You measure a set of standards, fit a line through them, and use the line to turn the signal of an unknown sample back into a concentration.

    How do you calculate a concentration from a standard curve?

    Fit the line y = mx + b through the standards, then rearrange it: x = (y − b) ÷ m. With a slope of 0.0495571 and an intercept of 0.00338095, a signal of 0.250 gives (0.250 − 0.00338095) ÷ 0.0495571 = 4.976. Multiply by the dilution factor if the sample was diluted.

    How many standards do I need?

    This calculator needs at least three. Five to eight standards spread evenly across the working range is the usual practice, because two points always give a perfect line and tell you nothing about linearity.

    What is a good R² for a standard curve?

    Many laboratory methods ask for R² of 0.99 or higher, and 0.995 or 0.999 for quantitative assays. The limit is set by your method or standard operating procedure, not by the statistics. A high R² alone does not prove the curve is linear, so look at the points against the line as well.

    Can I read a sample that is above the highest standard?

    You should not. The line is only supported between the lowest and highest standard, and many assays flatten at high concentration. Dilute the sample so its signal falls inside the range, measure again, and multiply the result by the dilution factor. The calculator flags any unknown outside the range.